Comparison of the sodium, potassium, chloride and glucose levels measured by a blood gas analyzer and an autoanalyzer

Main Article Content

Omer Kaya
Kamile Yucel

Abstract

Aim: A blood gas analyzers (BGA) are vital equipment frequently used in emergency departments and intensive care units. It is clinically important that the measurements of a BGA and an autoanalyzer (AA) provide equivalent results, which is confirmed by their proximitiy to the absolute value. This study aimed to compare the sodium (Na+), potassium (K+), chloride (Cl-) and glucose values in venous blood samples measured with a BGA and a standard AA with external quality control values.Material and Methods: The results of patients that presented to our emergency department between April 1, 2019 and July 1, 2019 and underwent the measurements of Na+ (n = 5,908), K+ (n = 5,755), Cl- (n = 5,101) and glucose (n = 5,871) simultaneously by BGA and AA were retrospectively compared. Results: In the Spearman correlation analysis between the two measurements, the correlation coefficient (r) was found as 0.78, 0.88, 0.89 and 0.97 for Na+, K+, Cl- and glucose, respectively. According to the Bland-Altman analysis, in the comparison of Na+, K+, Cl- and glucose values, the average bias percentages at the 95% confidence interval were -0.8 (4.8 to -6.4), -9 (8.6 to -26.5), -0.2 (6.2 to -6.5), and 0.3 (21.8 to -21.3), respectively. Conclusion: We concluded that the Na+ and Cl- results obtained from BGA can be used instead of those obtained from an AA; however, the blood gas K+ and glucose results of the former cannot replace those of the latter.

Downloads

Download data is not yet available.

Article Details

Section

Original Articles

How to Cite

1.
Comparison of the sodium, potassium, chloride and glucose levels measured by a blood gas analyzer and an autoanalyzer . Ann Med Res [Internet]. 2021 May 25 [cited 2026 Apr. 30];27(10):2639-44. Available from: http://www.annalsmedres.org/index.php/aomr/article/view/1004

References